An incomplete pacemaker lead fracture causing inappropriate pacing inhibition due to oversensing of the minute ventilation sensor pulses
Cardiac pacemakers using rate response technologies represent the effectiveness of increasing the heart rate and cardiac output during exercise. Minute ventilation (MV) sensors are popular and estimate rates using transthoracic impedance by emitting very low amperage short electrical current pulses...