Penetration of inferior vena cava filter strut through an introducer sheath inserted from the left internal jugular vein: An anatomical pitfall

The report describes a case of penetration of an inferior vena cava filter (IVCF) strut through an introducer sheath inserted from the left internal jugular vein. A 64-year-old woman was admitted to our hospital with pulmonary embolism and lower extremity deep vein thrombosis. We attempted to implan...