Catastrophic failure of tripolar constrained liners due to backside wear: a novel failure mode

Constrained acetabular liners have been developed for patients who are at high risk for dislocation or who are undergoing revision surgery for recurrent dislocations. We report on 2 cases of failure of tripolar constrained liners due to severe backside polyethylene wear after dissociation of the out...