Evaluation of prototype of improved electron collimation system for Elekta linear accelerators
PURPOSE: This study evaluated a new electron collimation system design for Elekta 6–20 MeV beams, which should reduce applicator weights by 25%–30%. Such reductions, as great as 3.9 kg for the largest applicator, should result in considerably easier handling by members of the radiotherapy team. METH...
Na minha lista:
| Publicado no: | J Appl Clin Med Phys |
|---|---|
| Main Authors: | , , |
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
John Wiley and Sons Inc.
2018
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6036346/ https://ncbi.nlm.nih.gov/pubmed/29756267 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/acm2.12342 |