Wear performance of cobalt chromium, ceramic, and oxidized zirconium on highly crosslinked polyethylene at mid-term follow-up

Patients with THA using a ceramic head and using an OxZr head were each matched to patients with a CoCr head. Mean implantation time was 5 years. There was no difference in steady state wear rate between the ceramic (0.066 ± 0.050 mm/year) and CoCr match groups (0.052 ± 0.041 mm/year), or between th...

Ful tanımlama

Kaydedildi:
Detaylı Bibliyografya
Yayımlandı:J Orthop
Asıl Yazarlar: Teeter, Matthew G., MacLean, Colin J., Somerville, Lyndsay E., Howard, James L., McCalden, Richard W., Lanting, Brent A., Vasarhelyi, Edward M.