Загрузка...
Going the distance: validation of Acuros and AAA at an extended SSD of 400 cm
Accurate dose calculation and treatment delivery is essential for total body irradiation (TBI). In an effort to verify the accuracy of TBI dose calculation at our institution, we evaluated both the Varian Eclipse AAA and Acuros algorithms to predict dose distributions at an extended source‐to‐surfac...
Сохранить в:
| Опубликовано в: : | J Appl Clin Med Phys |
|---|---|
| Главные авторы: | , , |
| Формат: | Artigo |
| Язык: | Inglês |
| Опубликовано: |
John Wiley and Sons Inc.
2016
|
| Предметы: | |
| Online-ссылка: | https://ncbi.nlm.nih.gov/pmc/articles/PMC5875551/ https://ncbi.nlm.nih.gov/pubmed/27074473 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1120/jacmp.v17i2.5913 |
| Метки: |
Добавить метку
Нет меток, Требуется 1-ая метка записи!
|