Dual‐axis rotational coronary angiography can reduce peak skin dose and scattered dose: a phantom study
The purpose of this study was to evaluate peak skin dose received by the patient and scattered dose to the operator during dual‐axis rotational coronary angiography (DARCA), and to compare with those of standard coronary angiography (SA). An anthropomorphic phantom was used to simulate a patient und...