Sensitivity of array detector measurements in determining shifts of MLC leaf positions
Using a MatriXX 2D ionization chamber array, we evaluated the detection sensitivity of systematically introduced MLC leaf positioning shifts to test whether the conventional IMRT QA method can be used for quality assurance of an MLC tracking algorithm. Because of finite special resolution, we first...