Wird geladen...
Relative dosimetry with an MR‐linac: Response of ion chambers, diamond, and diode detectors for off‐axis, depth dose, and output factor measurements
PURPOSE: The purpose of this study was to acquire beam data for an MR‐linac, with and without a 1.5 T magnetic field, by using a variety of commercially available detectors to assess their relative response in the magnetic field. The impact of the magnetic field on the measured dose distribution was...
Gespeichert in:
| Veröffentlicht in: | Med Phys |
|---|---|
| Hauptverfasser: | , , , , |
| Format: | Artigo |
| Sprache: | Inglês |
| Veröffentlicht: |
John Wiley and Sons Inc.
2017
|
| Schlagworte: | |
| Online Zugang: | https://ncbi.nlm.nih.gov/pmc/articles/PMC5807173/ https://ncbi.nlm.nih.gov/pubmed/29178457 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/mp.12699 |
| Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|