Načítá se...

Relative dosimetry with an MR‐linac: Response of ion chambers, diamond, and diode detectors for off‐axis, depth dose, and output factor measurements

PURPOSE: The purpose of this study was to acquire beam data for an MR‐linac, with and without a 1.5 T magnetic field, by using a variety of commercially available detectors to assess their relative response in the magnetic field. The impact of the magnetic field on the measured dose distribution was...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:Med Phys
Hlavní autoři: O'Brien, Daniel J., Dolan, James, Pencea, Stefan, Schupp, Nicholas, Sawakuchi, Gabriel O.
Médium: Artigo
Jazyk:Inglês
Vydáno: John Wiley and Sons Inc. 2017
Témata:
On-line přístup:https://ncbi.nlm.nih.gov/pmc/articles/PMC5807173/
https://ncbi.nlm.nih.gov/pubmed/29178457
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/mp.12699
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo otaguje tento záznam!