Carregant...

Verification of the accuracy of 3D calculations of breast dose during tangential irradiation: measurements in a breast phantom

This report specifically describes the use of a unique anthropomorphic breast phantom to validate the accuracy of three‐dimensional dose calculations performed by a commercial treatment‐planning system for intact‐breast tangential irradiation. The accuracy of monitor‐unit calculations has been corro...

Descripció completa

Guardat en:
Dades bibliogràfiques
Publicat a:J Appl Clin Med Phys
Autors principals: Kirsner, Steven M., Prado, Karl L., Tailor, Ramesh C., Bencomo, José A.
Format: Artigo
Idioma:Inglês
Publicat: John Wiley and Sons Inc. 2001
Matèries:
Accés en línia:https://ncbi.nlm.nih.gov/pmc/articles/PMC5726046/
https://ncbi.nlm.nih.gov/pubmed/11602011
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1120/jacmp.v2i3.2608
Etiquetes: Afegir etiqueta
Sense etiquetes, Sigues el primer a etiquetar aquest registre!