A carregar...
The overshoot phenomenon in step‐and‐shoot IMRT delivery
The control loop in the Varian DMLC system (V4.8) requires [Formula: see text] msec to monitor and halt the irradiation of a segment, causing an “overshoot” effect: the segment ends on a fractional monitor unit larger than that planned. As a result, the actual MU delivered may differ from that plann...
Na minha lista:
| Publicado no: | J Appl Clin Med Phys |
|---|---|
| Main Authors: | , |
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
John Wiley and Sons Inc.
2001
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC5726041/ https://ncbi.nlm.nih.gov/pubmed/11602010 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1120/jacmp.v2i3.2607 |
| Tags: |
Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
|