Electron energy constancy verification using a double‐wedge phantom
Routine constancy checks of electron energy are often time consuming because of the necessity to measure a dose at two depths. A technique is described that uses a double‐wedge shaped phantom positioned on a Profiler™ diode array for measuring an electron energy constancy metric similar to [Formula:...