Systematic offset of kV and MV localization systems as a function of gantry angle

Mechanical flex of the gantry and mounted imaging panels leads to systematic offsets in localization image isocenter as a function of gantry angle for linear accelerator‐mounted image guidance systems. Subsequently, object positions obtained from localization radiographs may be offset, resulting in...