A Monte Carlo and experimental investigation of the dosimetric behavior of low‐ and medium‐perturbation diodes used for entrance in vivo dosimetry in megavoltage photon beams
Full buildup diodes can cause significant dose perturbation if they are used on most or all of radiotherapy fractions. Given the importance of frequent in vivo measurements in complex treatments, using thin buildup (low‐perturbation) diodes instead is gathering interest. However, such diodes are str...