Sensitivity of a helical diode array device to delivery errors in IMRT treatment and establishment of tolerance level for pretreatment QA

The aim of this study is to determine the gantry angle and multileaf collimator (MLC) gap error‐detection threshold of a diode helical array with an inserted microionization chamber in order to use this device for the pretreatment quality assurance (QA) of intensity‐modulated radiation therapy (IMRT...

Descripción completa

Guardado en:
Detalles Bibliográficos
Publicado en:J Appl Clin Med Phys
Autores principales: García‐Vicente, Feliciano, Fernández, Virginia, Bermüdez, Rocío, Gómez, Alberto, Pérez, Leopoldo, Zapatero, Almudena, Torres, Juan J.
Formato: Artigo
Lenguaje:Inglês
Publicado: John Wiley and Sons Inc. 2012
Materias:
Acceso en línea:https://ncbi.nlm.nih.gov/pmc/articles/PMC5716126/
https://ncbi.nlm.nih.gov/pubmed/22231218
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1120/jacmp.v13i1.3660
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!