Development of a modified head and neck quality assurance phantom for use in stereotactic radiosurgery trials

An anthropomorphic head phantom, constructed from a water‐equivalent plastic shell with only a spherical target, was modified to include a nonspherical target (pituitary) and an adjacent organ at risk (OAR) (optic chiasm), within 2 mm, simulating the anatomy encountered when treating acromegaly. The...