Modeling the Agility MLC in the Monaco treatment planning system

We investigate the relationship between the various parameters in the Monaco MLC model and dose calculation accuracy for an Elekta Agility MLC. The vendor‐provided MLC modeling procedure — completed first with external vendor participation and then exclusively in‐house — was used in combination with...