Management of Elevated Therapeutic Impedances on Deep Brain Stimulation Leads
CLINICAL VIGNETTE: A 64-year-old male with a history of essential tremor with bilateral thalamic ventralis intermedius deep brain stimulation implants had elevated therapeutic impedance values despite normal lead integrity impedances and good response to stimulation. CLINICAL DILEMMA: Do elevated th...