Acute Shunt Failure Due to Perforation of Ventriculoperitoneal Shunt Tubing during Percutaneous Gastrostomy
The placement of a percutaneous endoscopic gastrostomy (PEG) in a patient with a pre-existing ventriculoperitoneal shunt is generally regarded as safe. A critical but often overlooked technical consideration is confirmation of the course of the distal shunt tube prior to PEG insertion. We present th...