Relation Between Wire Resistance and Fluid Pressure in the Transient Hot-Wire Method

The resistance of metals is a function of applied pressure, and this dependence is large enough to be significant in the calibration of transient hot-wire thermal conductivity instruments. We recommend that for the highest possible accuracy, the instrument’s hot wires should be calibrated in situ. I...