Endocardial pacing electrode design and rate of displacement.
In a series of 184 patients with a permanent endocardial pacing system the rate of electrode displacement was 3-0 per cent using the Devices L120SR electrode, and 23-1 per cent using the Devices S120 electrode. The technique and pacing team were unchanged during the period of study. It is suggested...
Na minha lista:
| Udgivet i: | Br Heart J |
|---|---|
| Principais autores: | , , , |
| Format: | Artigo |
| Sprog: | Inglês |
| Udgivet: |
BMJ Publishing Group
1977
|
| Fag: | |
| Online adgang: | https://ncbi.nlm.nih.govhttps://pmc.ncbi.nlm.nih.gov/articles/PMC483268/ https://ncbi.nlm.nih.govhttps://pubmed.ncbi.nlm.nih.gov/861094/ https://ncbi.nlm.nih.govhttps://doi.org/10.1136/hrt.39.5.515 |
| Tags: |
Ingen Tags, Vær først til at tagge denne postø!
|