An evaluation of the Celloscope 401 electronic blood cell counter

For counting erythrocytes the instrument was precise, with a mean coefficient of variation of 1·21%. Erythrocyte counts showed close agreement with results obtained on a Coulter A electronic counter of proven accuracy. When the Celloscope 401 was modified by the manufacturers to eliminate electrical...