High counting rates from (125)I in radioimmunoassays
Responses from one manual and four automatic gamma counters for assaying (125)I were compared. The resolution times quoted by manufacturers when applied in a standardized correction formula to obtain true counts from observed counting rates above 10(4) counts per second appeared to be inadequate. Th...