The effect of motion on IMRT – looking at interplay with 3D measurements

Six base of skull IMRT treatment plans were delivered to 3D dosimeters within the RPC Head and Neck Phantom for QA verification. Isotropic 2mm 3D data was obtained using the DLOS-PRESAGE system and compared to an Eclipse (Varian) treatment plan. Normalized Dose Distribution pass rates were obtained...

Täydet tiedot

Tallennettuna:
Bibliografiset tiedot
Julkaisussa:J Phys Conf Ser
Päätekijät: Thomas, A, Yan, H, Oldham, M, Juang, T, Adamovics, J, Yin, FF
Aineistotyyppi: Artigo
Kieli:Inglês
Julkaistu: 2013