Lanean...

Relationship between the physical size, incompetence, and stenosis of prosthetic mitral valves

One each of 17 commercially available prosthetic mitral valves has been subjected to in vitro testing using a pulse duplicator. Measurements of mean diastolic pressure difference, incompetence, dimensions, mechanical movements, and turbulence were made, and the quality of manufacture was examined. A...

Deskribapen osoa

Gorde:
Xehetasun bibliografikoak
Egile Nagusiak: Wright, J. T. M., Temple, L. J.
Formatua: Artigo
Hizkuntza:Inglês
Argitaratua: 1972
Gaiak:
Sarrera elektronikoa:https://ncbi.nlm.nih.gov/pmc/articles/PMC472586/
https://ncbi.nlm.nih.gov/pubmed/5039443