Chargement en cours...

A DXA Whole Body Composition Cross-Calibration Experience: Evaluation with Humans, Spine and Whole Body Phantoms

New densitometer installation requires cross-calibration for accurate longitudinal assessment. When replacing a unit with the same model, ISCD recommends cross-calibrating by scanning phantoms 10 times on each instrument and states spine BMD should be within 1%, while total body lean, fat and %fat m...

Description complète

Enregistré dans:
Détails bibliographiques
Publié dans:J Clin Densitom
Auteurs principaux: Krueger, D., Libber, J., Sanfilippo, J., Yu, H.J., Horvath, B., Miller, C.G., Binkley, N.
Format: Artigo
Langue:Inglês
Publié: 2015
Sujets:
Accès en ligne:https://ncbi.nlm.nih.gov/pmc/articles/PMC4674392/
https://ncbi.nlm.nih.gov/pubmed/26071169
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1016/j.jocd.2015.04.003
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!