A DXA Whole Body Composition Cross-Calibration Experience: Evaluation with Humans, Spine and Whole Body Phantoms
New densitometer installation requires cross-calibration for accurate longitudinal assessment. When replacing a unit with the same model, ISCD recommends cross-calibrating by scanning phantoms 10 times on each instrument and states spine BMD should be within 1%, while total body lean, fat and %fat m...