Evaluation of 5 µm Superficially Porous Particles for Capillary and Microfluidic LC Columns
Large-size (4–5 µm) superficially porous particles yield lower plate heights (e.g., the minimal reduced plate height or h(min) ≈ 1.5) than fully porous particles of a similar size when packed into large-bore columns. This property allows for better chromatographic performance without the higher pres...