Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry
A synthetic single crystal diamond detector (SCDD) is commercially available and is characterized for radiation dosimetry in various radiation beams in this study. The characteristics of the commercial SCDD model 60019 (PTW) with 6- and 15-MV photon beams, and 208-MeV proton beams, were investigated...
Kaydedildi:
| Yayımlandı: | J Radiat Res |
|---|---|
| Asıl Yazarlar: | , , , , |
| Materyal Türü: | Artigo |
| Dil: | Inglês |
| Baskı/Yayın Bilgisi: |
Oxford University Press
2015
|
| Konular: | |
| Online Erişim: | https://ncbi.nlm.nih.gov/pmc/articles/PMC4628217/ https://ncbi.nlm.nih.gov/pubmed/26268483 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1093/jrr/rrv044 |