Highly Sensitive Amperometric Pt-Nafion Gas Phase Nitric Oxide Sensor: Performance and Application in Characterizing Nitric Oxide-Releasing Biomaterials

A highly sensitive amperometric gas-phase nitric oxide (NO) sensor based on a Pt working electrode chemically deposited on a Nafion film is described. The Pt electrode is chemically deposited on a Nafion 117 membrane by impregnating the film with Pt(NH(3))(4)(2+) ions, which are then exposed to NaBH...