PURPOSE: To evaluate the effect of attenuation of MR coils on quantitative carotid PET/MR exams. Additionally, to develop and evaluate automated attenuation correction method for flexible carotid MR coils was developed and evaluated. METHODS: The attenuation of the carotid coil was measured by imagi...