A New Case of Fracture of a Modular Femoral Neck Device After a Total Hip Arthroplasty

This report addresses a new case of a modular femoral neck fracture after total hip arthroplasty. A now seventy-three-year- old overweight female underwent total hip replacement of the right hip because of osteoarthritis in the year 2002. Then, years later, the head and the polyethylene were changed...