Passively scattered proton beam entrance dosimetry with a plastic scintillation detector

We tested the feasibility of using plastic scintillation detectors (PSDs) for proton entrance dosimetry. A PSD built with BCF-12 scintillating fiber was used to measure the absolute entrance dose of a passively scattered proton beam for energies ranging from 140 MeV to 250 MeV, and for a range of sp...