Development and Optimization of a Real-Time PCR Assay for Detection of Herpes Simplex and Varicella-Zoster Viruses in Skin and Mucosal Lesions by Use of the BD Max Open System

We transitioned laboratory-developed PCR assays for herpes simplex virus 1 (HSV-1), HSV-2, and varicella-zoster virus (VZV) to the BD Max system by using BD Max open system reagents. After optimization, the agreement with the reference PCR assay was 100% (123/123) for HSV-1, 96.7% (119/123) for HSV-...