Bubble formation occurs in insulin pumps in response to changes in ambient temperature and atmospheric pressure but not as a result of vibration
INTRODUCTION: Bubble formation in insulin pump giving sets is a common problem. We studied change in temperature, change in atmospheric pressure, and vibration as potential mechanisms of bubble formation. METHODS: 5 Animas 2020 pumps with 2 mL cartridges and Inset II infusion systems, 5 Medtronic Pa...