Bubble formation occurs in insulin pumps in response to changes in ambient temperature and atmospheric pressure but not as a result of vibration

INTRODUCTION: Bubble formation in insulin pump giving sets is a common problem. We studied change in temperature, change in atmospheric pressure, and vibration as potential mechanisms of bubble formation. METHODS: 5 Animas 2020 pumps with 2 mL cartridges and Inset II infusion systems, 5 Medtronic Pa...

Description complète

Enregistré dans:
Détails bibliographiques
Auteurs principaux: Lopez, Prudence E, King, Bruce R, Goss, Peter W, Chockalingam, Ganesh
Format: Artigo
Langue:Inglês
Publié: BMJ Publishing Group 2014
Sujets:
Accès en ligne:https://ncbi.nlm.nih.gov/pmc/articles/PMC4212562/
https://ncbi.nlm.nih.gov/pubmed/25452873
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1136/bmjdrc-2014-000036