A comparison between two NIRS oximeters (INVOS, OxyPrem) using measurement on the arm of adults and head of infants after caesarean section
Previously the NIRS oximeter OxyPrem was calibrated by comparison to the INVOS in a blood-lipid phantom. The aim of the present study was to test this calibration clinically. During vasculur occlusions in 10 adults and after birth in 25 term infants the relationship was OxyPrem = 1.24 x INVOS - 23.6...