Discrepancy in Optical & Mechanical Method in Coagulation Tests in a Turbid Sample
Automated coagulation analyzers have replaced manual methods to meet the ever-increasing test load in many laboratories. Of the different methods, two distinct methods exist based on optical and mechanical clot detection (magnetic steel ball method). In optical method the detection of clot formation...