Dose error from deviation of dwell time and source position for high dose-rate (192)Ir in remote afterloading system

The influence of deviations in dwell times and source positions for (192)Ir HDR-RALS was investigated. The potential dose errors for various kinds of brachytherapy procedures were evaluated. The deviations of dwell time ΔT of a (192)Ir HDR source for the various dwell times were measured with a well...