Cargando...

Dose error from deviation of dwell time and source position for high dose-rate (192)Ir in remote afterloading system

The influence of deviations in dwell times and source positions for (192)Ir HDR-RALS was investigated. The potential dose errors for various kinds of brachytherapy procedures were evaluated. The deviations of dwell time ΔT of a (192)Ir HDR source for the various dwell times were measured with a well...

Descrición completa

Gardado en:
Detalles Bibliográficos
Main Authors: Okamoto, Hiroyuki, Aikawa, Ako, Wakita, Akihisa, Yoshio, Kotaro, Murakami, Naoya, Nakamura, Satoshi, Hamada, Minoru, Abe, Yoshihisa, Itami, Jun
Formato: Artigo
Idioma:Inglês
Publicado: Oxford University Press 2014
Assuntos:
Acceso en liña:https://ncbi.nlm.nih.gov/pmc/articles/PMC4099994/
https://ncbi.nlm.nih.gov/pubmed/24566719
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1093/jrr/rru001
Tags: Engadir etiqueta
Sen Etiquetas, Sexa o primeiro en etiquetar este rexistro!