Caricamento...

Evaluation of imaging performance of megavoltage cone-beam CT over an extended period

A linear accelerator vendor and the AAPM TG-142 report propose that quality assurance testing for image-guided devices such megavoltage cone-beam CT (MV-CBCT) be conducted on a monthly basis. In clinical settings, however, unpredictable errors such as image artifacts can occur even when quality assu...

Descrizione completa

Salvato in:
Dettagli Bibliografici
Autori principali: Sumida, Iori, Yamaguchi, Hajime, Kizaki, Hisao, Yamada, Yuji, Koizumi, Masahiko, Yoshioka, Yasuo, Ogawa, Kazuhiko, Kakimoto, Naoya, Murakami, Shumei, Furukawa, Souhei
Natura: Artigo
Lingua:Inglês
Pubblicazione: Oxford University Press 2014
Soggetti:
Accesso online:https://ncbi.nlm.nih.gov/pmc/articles/PMC3885132/
https://ncbi.nlm.nih.gov/pubmed/23979076
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1093/jrr/rrt100
Tags: Aggiungi Tag
Nessun Tag, puoi essere il primo ad aggiungerne! !