Načítá se...

Non-obvious Problems in Clark Electrode Application at Elevated Temperature and Ways of Their Elimination

Well-known cause of frequent failures of closed oxygen sensors is the appearance of gas bubbles in the electrolyte. The problem is traditionally associated with insufficient sealing of the sensor that is not always true. Study of a typical temperature regime of measurement system based on Clark sens...

Celý popis

Uloženo v:
Podrobná bibliografie
Hlavní autoři: Miniaev, M. V., Belyakova, M. B., Kostiuk, N. V., Leshchenko, D. V., Fedotova, T. A.
Médium: Artigo
Jazyk:Inglês
Vydáno: Hindawi Publishing Corporation 2013
Témata:
On-line přístup:https://ncbi.nlm.nih.gov/pmc/articles/PMC3747604/
https://ncbi.nlm.nih.gov/pubmed/23984188
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1155/2013/249752
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo otaguje tento záznam!