Učitavanje...

Concept of defibrillation vector in the management of high defibrillation threshold

We present a case where defibrillation threshold was dangerously elevated to the point that the patient had no safety margin, and his implantable cardioverter-defibrillator generator was discovered to have migrated. Generator migration reduces the distance between the can and the coil, effectively c...

Cijeli opis

Spremljeno u:
Bibliografski detalji
Glavni autori: Hayes, Kevin, Deshmukh, Abhishek, Pant, Sadip, Tobler, Gareth, Paydak, Hakan
Format: Artigo
Jezik:Inglês
Izdano: Baishideng Publishing Group Co., Limited 2013
Teme:
Online pristup:https://ncbi.nlm.nih.gov/pmc/articles/PMC3653010/
https://ncbi.nlm.nih.gov/pubmed/23675557
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.4330/wjc.v5.i4.106
Oznake: Dodaj oznaku
Bez oznaka, Budi prvi tko označuje ovaj zapis!