Haemothorax following bullectomy caused by a sharp edge of the Endoloop

A 22-year old man presented with a massive haemothorax 25 days after bullectomy for a spontaneous pneumothorax. Thoracoscopic surgery revealed ongoing bleeding from the chest wall caused by a sharp edge of the Endoloop Ligature (Ethicon Endo-Surgery) used to resect the remaining small part of the lu...