Stress fracture of the femoral component in total knee replacement: a report of 3 cases

We report 3 cases of stress fracture of the femoral component (small size Whiteside Ortholoc II) in total knee replacements which occurred at 32, 52, and 73 months after operation. The site was at the junction between the medial posterior bevelled surface and the posterior flange. The most likely ca...