Flow-through comb electroporation device for delivery of macromolecules

We present a microfluidic electroporation device with a comb electrode layout fabricated in polydimethylsiloxane (PMDS) and glass. Characterization experiments with HeLa cells and fluorescent dextran show efficient delivery (~95%) with low toxicity (cell viability ~85%) as well as rapid pore closure...