Wordt geladen...

Safety and efficacy of the Aperio thrombectomy device when compared to the Solitaire AB/FR and the Revive devices in a pulsatile flow system

Background and Purpose: There are a limited number of studies comparing the Aperio mechanical thrombectomy device to other stent-based devices. In this paper, we compared the Aperio thrombectomy device to the Solitaire AB, FR and Revive devices in a model of the middle cerebral artery (MCA) within a...

Volledige beschrijving

Bewaard in:
Bibliografische gegevens
Hoofdauteurs: Saleh, Mahdi, Spence, John Nathan, Nayak, Sanjeev, Pearce, Gillian, Tennuci, Christopher, Roffe, Christine
Formaat: Artigo
Taal:Inglês
Gepubliceerd in: e-Century Publishing Corporation 2012
Onderwerpen:
Online toegang:https://ncbi.nlm.nih.gov/pmc/articles/PMC3499937/
https://ncbi.nlm.nih.gov/pubmed/23173104