Positioner and Clothing Artifact Can Affect One-third Radius BMD Measurement

This report identifies a radius DXA confounder and technical approach to avoid this inaccuracy. Initially, a precision study revealed substantial differences (p < 0.001) in radius BMD least significant change (LSC) ranging from 0.038 to 0.073 g/cm(2) between three technologists that each performe...