A carregar...
Characterization of a high resolution hybrid DOI detector for a dedicated breast PET/CT scanner
The aim of this study is to design and test a new high resolution hybrid depth of interaction (DOI) detector for a dedicated breast PET/CT scanner. Two detectors have been designed and built. The completed detectors are based on a 14×14 array of 1.5×1.5×20 mm(3)unpolished lutetium orthosilicate (LSO...
Na minha lista:
| Main Authors: | , , , , |
|---|---|
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
2012
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC3397825/ https://ncbi.nlm.nih.gov/pubmed/22581109 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1088/0031-9155/57/11/3435 |
| Tags: |
Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
|