Characterization of a high resolution hybrid DOI detector for a dedicated breast PET/CT scanner
The aim of this study is to design and test a new high resolution hybrid depth of interaction (DOI) detector for a dedicated breast PET/CT scanner. Two detectors have been designed and built. The completed detectors are based on a 14×14 array of 1.5×1.5×20 mm(3)unpolished lutetium orthosilicate (LSO...