Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images

Purpose: To evaluate the geometric accuracy of beam targeting in external surrogate-based gated volumetric modulated arc therapy (VMAT) using kilovoltage (kV) x-ray images acquired during dose delivery. Methods: Gated VMAT treatments were delivered using a Varian TrueBeam STx Linac for both physical...

Täydet tiedot

Tallennettuna:
Bibliografiset tiedot
Päätekijät: Li, Ruijiang, Mok, Edward, Han, Bin, Koong, Albert, Xing, Lei
Aineistotyyppi: Artigo
Kieli:Inglês
Julkaistu: American Association of Physicists in Medicine 2012
Aiheet:
Linkit:https://ncbi.nlm.nih.gov/pmc/articles/PMC3344884/
https://ncbi.nlm.nih.gov/pubmed/22559639
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1118/1.4704729