Characterization of a high-purity germanium detector for small-animal SPECT

We present an initial evaluation of a mechanically-cooled, high-purity germanium double-sided strip detector as a potential gamma camera for small-animal SPECT. It is 90 mm in diameter and 10 mm thick with two sets of 16 orthogonal strips that have a 4.5 mm width with a 5 mm pitch. We found an energ...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Main Authors: Johnson, Lindsay C, Campbell, Desmond L, Hull, Ethan L, Peterson, Todd E
Formato: Artigo
Idioma:Inglês
Publicado em: 2011
Assuntos:
Acesso em linha:https://ncbi.nlm.nih.gov/pmc/articles/PMC3178965/
https://ncbi.nlm.nih.gov/pubmed/21852723
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1088/0031-9155/56/18/007