A Simple Method for Measuring Organotypic Tissue Slice Culture Thickness
This paper presents a simple method to measure tissue slice thicknesses using an ohmmeter. The circuit described here is composed of a metal probe, an ohmmeter, a counter electrode, culture medium or physiological buffer, and tissue slice. The probe and the electrode are on opposite interfaces of an...